Combined magnetic resonance imaging and ultrasound echography guidance for motion compensated HIFU interventions

Mario Ries, Baudouin Denis De Senneville, Yvan Regard, Chrit Moonen

Onderzoeksoutput: Hoofdstuk in Boek/Rapport/CongresprocedureConferentiebijdragepeer review

7 Citaten (Scopus)

Samenvatting

The objective of this study is to evaluate the feasibility to integrate ultrasound echography as an additional imaging modality for continuous target tracking, while performing simultaneously real-time MR- thermometry to guide a High Intensity Focused Ultrasound (HIFU) ablation. Experiments on a moving phantom were performed with MRI-guided HIFU during continuous ultrasound echography. Real-time US echography-based target tracking during MR-guided HIFU heating was performed with heated area dimensions similar to those obtained for a static target. The combination of both imaging modalities shows great potential for real-time beam steering and MR-thermometry.

Originele taal-2Engels
Titel12th International Symposium on Therapeutic Ultrasound
Pagina's202-206
Aantal pagina's5
DOI's
StatusGepubliceerd - 2012
Extern gepubliceerdJa
Evenement12th International Symposium on Therapeutic Ultrasound, ISTU 2012 - Heidelberg, Duitsland
Duur: 10 jun. 201213 jun. 2012

Publicatie series

NaamAIP Conference Proceedings
Volume1503
ISSN van geprinte versie0094-243X
ISSN van elektronische versie1551-7616

Congres

Congres12th International Symposium on Therapeutic Ultrasound, ISTU 2012
Land/RegioDuitsland
StadHeidelberg
Periode10/06/1213/06/12

Vingerafdruk

Duik in de onderzoeksthema's van 'Combined magnetic resonance imaging and ultrasound echography guidance for motion compensated HIFU interventions'. Samen vormen ze een unieke vingerafdruk.

Citeer dit